The self-expanding Symetis Acurate does not increase cerebral microembolic load when compared to the balloon-expandable Edwards Sapien prosthesis: a transcranial Doppler study in patients undergoing transapical aortic valve implantation

PLoS One. 2014 Oct 7;9(10):e108191. doi: 10.1371/journal.pone.0108191. eCollection 2014.

Abstract

Objectives: The aim of this study was to quantify potential differences in count, frequency and pattern of high-intensity transient signals (HITS) during transapical transcatheter aortic valve implantation (TA-TAVI), by comparing the Symetis Acurate TA (SA) with the balloon-expandable Edwards Sapien XT (ES) system.

Background: Recently, the Symetis Acurate TA revalving system has been introduced for TA-TAVI. The Symetis Acurate TA aortic bioprosthesis is self-expanding and is deployed by a specific two-step implantation technique. Whether this novel method increases the load of intraprocedural emboli, detected by transcranial Doppler ultrasound (TCD) as HITS, or not is not clear.

Methods: Twenty-two patients (n = 11 in each study arm, median logistic EuroScore 20%, median STS score 7%) displayed continuous TCD signals of good quality throughout the entire TA-TAVI procedure and were included in the final analysis. Data are presented as median with interquartile ranges.

Results: No significant differences were detected in total procedural or interval-related HITS load (SA: 303 [200; 594], ES: 499 [285; 941]; p = 0.16). With both devices, HITS peaked during prosthesis deployment (PD), whereas significantly fewer HITS occurred during instrumentation (SA: p = 0.002; ES: < 0.001) or post-implantation PI (SA: p = 0.007; ES: < 0.001). PD-associated HITS amounted to almost half of the total HITS load. One patient suffered new disabling stroke at 30 days. Thirty-day mortality amounted to 13.6% (3 of 22 patients).

Conclusions: Simplified transapical delivery using the self-expanding SA device does not increase HITS, despite of a two-step deployment technique with more interactions with the native aortic valve, when compared to the balloon-expandable ES valve. The similarity in HITS count, frequency and pattern with the two systems suggests a common mechanism for the release of cerebral microemboli.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects*
  • Heart Valve Prosthesis Implantation / methods*
  • Humans
  • Intracranial Embolism / diagnosis*
  • Intracranial Embolism / etiology*
  • Male
  • Time Factors
  • Ultrasonography, Doppler, Transcranial* / methods

Grants and funding

Department of Anaesthesiology and Pain Therapy, University Hospital Bern, provided support for the maintenance of the emboli detection machine (Transcranial Doppler Sonography). This funder did not have any role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.