Contained Rupture in HeartMate II Outflow Graft Bend Relief Disconnection

Asian Cardiovasc Thorac Ann. 2016 Mar;24(3):269-71. doi: 10.1177/0218492314557871. Epub 2014 Oct 27.

Abstract

Thoratec recalled their HeartMate II ventricular assist device in March 2012 after some problems related to disconnection of the bend relief. Abrasion on the pump outflow graft caused by a disconnected bend relief is rare. We report the case of a 49-year-old man in whom a disconnected bend relief caused a puncture in the outflow graft, resulting in a contained hematoma, requiring outflow graft replacement.

Keywords: Heart failure; heart-assist devices; prosthesis failure; punctures; thoracic surgical procedures.