Validation of the Pangao PG-800A11 wrist device assessed according to the European Society of Hypertension and the British Hypertension Society protocols

Blood Press Monit. 2015 Apr;20(2):108-11. doi: 10.1097/MBP.0000000000000095.

Abstract

Aims: The objective of this study was to determine the accuracy of the Pangao PG-800A11 wrist blood pressure monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010 and the protocol of the British Hypertension Society (BHS).

Methods: The device evaluations were performed in 85 participants, 33 of whom were included according to the ESH-IP revision 2010 and 52 of whom were included on the basis of the requirements of the BHS protocol. The validation procedure and data analysis followed the protocols precisely.

Results: The device achieved an A/A grading for the BHS protocol and passed all phases of the ESH-IP revision 2010 protocol. The mean difference ±SD for the ESH and BHS protocols, respectively, was -0.6±4.5 and -0.8±6.2 mmHg for systolic pressure and 1.2±4.6 and -0.5±5.1 mmHg for diastolic pressure. The device maintained its A/A grading throughout the low, medium, and high-pressure ranges.

Conclusion: The Pangao PG-800A11 wrist blood pressure monitor passed all requirements of the ESH-IP revision 2010 and achieved A/A grade of the BHS protocol in an adult population.