Purpose: Minimally invasive spine stabilization (MISt) procedures, including MIS-transforaminal lumbar interbody fusion (MIS-TLIF), rely on precise placement of percutaneous pedicle screws (PPS). Serious intraoperative complications associated with PPS placement include great vessel and bowel injuries due to the guide-wire's anterior migration and penetration through the anterior aspect of the vertebral body. To address this issue, we developed a novel percutaneous guide wire (S-wire) and compared the biomechanical characteristics of S-wire and conventional wire in cadaveric spines, and to evaluate the S-wire's efficacy and safety in a clinical trial.
Methods: The S-wire is hollow, with braided wires extending at one tip. We compared the push-out and penetration forces of the S-wire and conventional wire in fresh cadaveric lumbar spines, from L1 to L5.
Results: Push-out forces caused the braided tip of the S-wire to bend or spread, and thus to resist anterior migration. The mean push-out forces for the S-wire and conventional wire were 15.5 ± 1.9 and 5.7 ± 0.8 N, respectively (P < .0001); the mean penetration forces were 69.1 ± 4.2 and 37.1± 4.8 N, respectively (P < .0005). There was no wire breakage or anterior-wall penetration in a clinical trial of 922 S-wires; interestingly, the pull-out force increased in 780 (84.6%) S-wires after placement.
Conclusions: The mean push-out and penetration forces for the S-wire were approximately 3 and 2 times greater than those of conventional wire, respectively. The S-wire effectively prevented guide-wire anterior migration and penetration of the anterior vertebral-body wall. The S-wire device should effectively improve the safety of MISt procedures, including MIS-TLIF and percutaneous kyphoplasty in selected patient with osteoporosis.
Keywords: Image-guided surgery; neurosurgery; spine surgery.
© The Author(s) 2014.