Non-invasive measurement of cardiac output using an iterative, respiration-based method

Br J Anaesth. 2015 Mar;114(3):406-13. doi: 10.1093/bja/aeu377. Epub 2014 Dec 8.

Abstract

Background: Current non-invasive respiratory-based methods of measuring cardiac output [Formula: see text] make doubtful assumptions and encounter significant technical difficulties. We present a new method using an iterative approach [Formula: see text], which overcomes limitations of previous methods.

Methods: Sequential gas delivery (SGD) is used to control alveolar ventilation [Formula: see text] and CO2 elimination [Formula: see text] during a continuous series of iterative tests. Each test consists of four breaths where inspired CO2 [Formula: see text] is controlled; raising end-tidal Pco2 [Formula: see text] by about 1.33 kPa (10 mm Hg) for the first breath, and then maintaining [Formula: see text] constant for the next three breaths. The [Formula: see text] required to maintain [Formula: see text] constant is calculated using the differential Fick equation (DFE), where [Formula: see text] is the only unknown and is arbitrarily assumed for the first iteration. Each subsequent iteration generates measures used for calculating [Formula: see text] by the DFE, refining the assumption of [Formula: see text] for the next test and converging it to the true [Formula: see text] when [Formula: see text] remains constant during the four test breaths. We compared [Formula: see text] with [Formula: see text] measured by bolus pulmonary artery thermodilution [Formula: see text] in seven pigs undergoing liver transplantation.

Results: [Formula: see text] implementation and analysis was fully automated, and [Formula: see text] varied from 0.6 to 5.4 litre min(-1) through the experiments. The bias (between [Formula: see text] and [Formula: see text]) was 0.2 litre min(-1) with 95% limit of agreement from -1.1 to 0.7 litre min(-1) and percentage of error of 32%. During acute changes of [Formula: see text], convergence of [Formula: see text] to actual [Formula: see text] required only three subsequent iterations.

Conclusions: [Formula: see text] measurement is capable of providing an automated semi-continuous non-invasive measure of [Formula: see text].

Keywords: carbon dioxide; cardiac output; monitoring; respiration.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Blood Gas Analysis / methods
  • Breath Tests / methods
  • Carbon Dioxide / metabolism
  • Cardiac Output / physiology*
  • Models, Animal
  • Monitoring, Physiologic / methods*
  • Oxygen Consumption / physiology
  • Pulmonary Gas Exchange / physiology
  • Reproducibility of Results
  • Respiration*
  • Swine
  • Thermodilution / methods
  • Tidal Volume / physiology

Substances

  • Carbon Dioxide