Volume and landmark analysis: comparison of MRI measurements obtained with an endorectal coil and with a phased-array coil

Clin Radiol. 2015 Apr;70(4):379-86. doi: 10.1016/j.crad.2014.12.002. Epub 2014 Dec 29.

Abstract

Aim: To compare prostate volumes and distances between anatomical landmarks on MRI images obtained with a phased-array coil (PAC) only and with a PAC and an endorectal coil (ERC).

Materials and methods: Informed consent was waived for this Health Insurance Portability and Accountability Act-compliant study. Fifty-nine men underwent PAC-MRI and ERC-MRI at 1.5 (n = 3) or 3 T (n = 56). On MRI images, two radiologists independently measured prostate volume and distances between the anterior rectal wall (ARW) and symphysis pubis at the level of the verumontanum; ARW and symphysis pubis at the level of the mid-symphysis pubis; and bladder neck and mid-symphysis pubis. Differences between measurements from PAC-MRI and ERC-MRI were assessed with the Wilcoxon RANK SUM test. Inter-reader agreement was assessed using the concordance correlation coefficient (CCC).

Results: Differences in prostate volume between PAC-MRI and ERC-MRI [median: -0.75 mm(3) (p = 0.10) and median: -0.84 mm(3) (p = 0.06) for readers 1 and 2, respectively] were not significant. For readers 1 and 2, median differences between distances were as follows: -10.20 and -12.75 mm, respectively, ARW to symphysis pubis at the level of the verumontanum; -6.60 and -6.08 mm, respectively, ARW to symphysis pubis at the level of the mid-symphysis pubis; -3 and -3 mm respectively, bladder neck to mid-symphysis pubis. All differences in distance were significant for both readers (p ≤ 0.0005). Distances were larger on PAC-MRI (p ≤ 0.0005). Inter-reader agreement regarding prostate volume was almost perfect on PAC-MRI (CCC: 0.99; 95% CI: 0.98-1.00) and ERC-MRI (CCC: 0.99; 95% CI: 0.99-1.00); inter-reader agreement for distance measurements varied (CCCs: 0.54-0.86).

Conclusion: Measurements of distances between anatomical landmarks differed significantly between ERC-MRI and PAC-MRI, although prostate volume measurements did not.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Humans
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Prostate / pathology*
  • Prostatic Neoplasms / diagnosis*
  • Reproducibility of Results
  • Retrospective Studies
  • Sensitivity and Specificity