Note: Position dependence of time signals picked off a microchannel plate detector

Rev Sci Instrum. 2015 Jan;86(1):016111. doi: 10.1063/1.4906327.

Abstract

Using an ultrafast laser and a precision mask, we demonstrate that time signals picked off directly from a microchannel plate detector depend on the position of the hit. This causes a time spread of about 280 ps, which can affect the quality of imaging measurements using large detectors.