A simple high-performance liquid chromatography for the determination of linezolid in human plasma and saliva

Biomed Chromatogr. 2015 Sep;29(9):1428-31. doi: 10.1002/bmc.3441. Epub 2015 Feb 10.

Abstract

Linezolid is an antimicrobial agent for the treatment of multiresistant Gram-positive infections. A practical high-performance liquid chromatography method was developed for the determination of linezolid in human plasma and saliva. Linezolid and an internal standard (o-ethoxybenzamide) were extracted from plasma and saliva with ethyl acetate and analyzed on a Capcell Pak C18 MG column with UV detection at 254 nm. The calibration curve was linear through the range 0.5-50 µg/mL using a 200 μL sample volume. The intra- and interday precisions were all <6.44% for plasma and 5.60% for saliva. The accuracies ranged from 98.8 to 110% for both matrices. The mean recoveries of linezolid were 80.8% for plasma and 79.0% for saliva. This method was used to determine the plasma and saliva concentrations of linezolid in healthy volunteers who were orally administered a 600 mg dose of linezolid. Our liquid-liquid extraction procedure is easy and requires a small volume of plasma or saliva (200 μL). This small volume can be advantageous in clinical pharmacokinetic studies, especially if children participate.

Keywords: HPLC-UV; linezolid; plasma; saliva; therapeutic drug monitoring.

Publication types

  • Evaluation Study

MeSH terms

  • Adult
  • Anti-Bacterial Agents / analysis*
  • Anti-Bacterial Agents / blood
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Linezolid / analysis*
  • Linezolid / blood
  • Male
  • Middle Aged
  • Saliva / chemistry*

Substances

  • Anti-Bacterial Agents
  • Linezolid