Automated electronic monitoring of circuit pressures during continuous renal replacement therapy: a technical report

Crit Care Resusc. 2015 Mar;17(1):51-4.

Abstract

Background: Automated electronic monitoring and analysis of circuit pressures during continuous renal replacement therapy (CRRT) has the potential to predict failure and allow intervention to optimise function.

Methods: Current CRRT machines can measure and store pressure readings for downloading into databases and for analysis. We developed a procedure to obtain such data at intervals of 1 minute and analyse them using the Prismaflex CRRT machine, and we present an example of such analysis.

Results: We obtained data on pressures obtained at intervals of 1 minute in a patient with acute kidney injury and sepsis treated with continuous haemofiltration at 2 L/hour of ultrafiltration and a blood flow of 200 mL/minute. Data analysis identified progressive increases in transmembrane pressure (TMP) and prefilter pressure (PFP) from time 0 until 33 hours or clotting. TMP increased from 104 mmHg to 313 mmHg and PFP increased from from 131 mmHg to 185 mmHg. Effluent pressure showed a progressive increase in the negative pressure applied to achieve ultrafiltration from 0 mmHg to -168 mmHg. The inflection point for such changes was also identified. Blood pathway pressures for access and return remained unchanged throughout.

Conclusions: Automated electronic monitoring of circuit pressure during CRRT is possible and provides useful information on the evolution of circuit clotting.

MeSH terms

  • Automation
  • Databases as Topic
  • Electronics, Medical*
  • Equipment Failure
  • Humans
  • Pressure*
  • Renal Replacement Therapy*