Energy response of GR-200A thermoluminescence dosemeters to 60Co and to monoenergetic synchrotron radiation in the energy range 28-40 keV

Radiat Prot Dosimetry. 2016 Jan;168(1):40-5. doi: 10.1093/rpd/ncv014. Epub 2015 Mar 2.

Abstract

The response of LiF:Mg,Cu,P thermoluminescence dosemeters (type GR-200A) to monoenergetic radiation of energy 28, 35, 38 and 40 keV was evaluated with respect to irradiation with a calibrated (60)Co gamma-ray source. High-precision measurements of the relative air kerma response performed at the SYRMEP beamline of the ELETTRA synchrotron radiation facility (Trieste, Italy) showed a significant deviation of the average response to low-energy X-rays from that to (60)Co, with an over-response from 6 % (at 28 keV) to 22 % (at 40 keV). These data are not consistent with literature data for these dosemeters, where model predictions gave deviation from unity of the relative air kerma response of about 10 %. The authors conclude for the need of additional determinations of the low-energy relative response of GR-200A dosemeters, covering a wider range of monoenergetic energies sampled at a fine energy step, as planned in future experiments by their group at the ELETTRA facility.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Air
  • Calibration
  • Cobalt Radioisotopes / analysis
  • Copper / chemistry
  • Fluorides / chemistry
  • Gamma Rays
  • Lithium Compounds / chemistry
  • Magnesium / chemistry
  • Phosphorus / chemistry
  • Photons
  • Radiation Dosage
  • Synchrotrons*
  • Thermoluminescent Dosimetry / methods*
  • X-Rays

Substances

  • Cobalt Radioisotopes
  • Lithium Compounds
  • lithium fluoride
  • Phosphorus
  • Copper
  • Magnesium
  • Fluorides