Radiofrequency in situ fenestration for aortic arch vessels during thoracic endovascular repair

J Endovasc Ther. 2015 Feb;22(1):116-21. doi: 10.1177/1526602814565776.

Abstract

Purpose: To report the first clinical application of a novel technique using radiofrequency puncture to create retrograde in situ fenestrations during thoracic endovascular aortic repair (TEVAR).

Methods: Between June 2011 and December 2013, 40 TEVAR procedures were performed in our facility, including 10 cases in which in situ fenestration was planned. Two thoracic stent-graft models were deployed: the Valiant (n=5) and the Zenith TX2 (n=5). A 0.035-inch PowerWire radiofrequency guidewire delivered from a brachial approach was used to fenestrate the grafts covering a left subclavian artery (LSA) in 9 cases and a left common carotid artery in one. The fenestrations were serially dilated to 6 mm, and self-expanding Advanta V12 covered stents were positioned in the target arteries.

Results: Technical success was achieved in 6 of the 10 planned cases. Of the remaining 4 cases, stent-grafts were deployed in zone 3 in 2 cases (one received a chimney to the LSA). Another stent-graft was deployed in zone 2 without endoleak after fenestration was abandoned (the LSA had good filling via the vertebral artery). In the last case, the fenestration was unsuccessful in double-layered (proximal extension overlap) stent-grafts; a carotid-axillary bypass was required. There were no fenestration-related complications, but overall surgical complications included a case of paraparesis that resolved following spinal drainage and a death from a preexisting aortoesophageal fistula. There were no postoperative strokes. All fenestrations remained patent, and there were no endoleaks at a mean 12-month follow-up (range 1-33).

Conclusion: Radiofrequency puncture is a viable alternative to needle or laser punctures for in situ fenestration during TEVAR. Early clinical results suggest technical feasibility and acceptable early outcomes.

Keywords: aortic arch; common carotid artery; covered stent; cutting balloon; in situ fenestration; left subclavian artery; radiofrequency; self-expanding stent-graft; thoracic aneurysm; thoracic endovascular aortic repair.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Aneurysm, Thoracic / mortality
  • Aortic Aneurysm, Thoracic / surgery*
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation*
  • Carotid Artery, Common / surgery*
  • Drug-Eluting Stents*
  • Endovascular Procedures* / adverse effects
  • Endovascular Procedures* / methods
  • Feasibility Studies
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Radio Waves
  • Retrospective Studies
  • Risk Factors
  • Subclavian Artery / surgery*
  • Treatment Outcome