Evaluation of a new double-lumen endobronchial tube with an integrated camera (VivaSight-DL(™) ): a prospective multicentre observational study

Anaesthesia. 2015 Aug;70(8):962-8. doi: 10.1111/anae.13068. Epub 2015 Apr 1.

Abstract

The VivaSight-DL(™) is a new single-use double-lumen endobronchial tube with an integrated camera. We studied this device in 151 consecutive patients scheduled for elective thoracic surgery in four different hospitals. Endobronchial intubation was successful in 148 patients (98%) (95% CI 94-99%). Median (IQR [range]) endobronchial intubation time was 59 (47-82 [17-932]) s and lung isolation was successfully achieved in 147 (99%) patients (95% CI 96-99%). A fibreoptic bronchoscope was required to assist endobronchial tube placement in 19 (13%) patients (95% CI 8-19%). Sore throat was reported by 37 (25%) patients (95% CI 18-33%), but no major complications were observed. We have reported the successful use of the VivaSight double-lumen tube for endobronchial intubation in a multicentre observational trial.

Publication types

  • Evaluation Study
  • Multicenter Study
  • Observational Study

MeSH terms

  • Aged
  • Equipment Design
  • Female
  • Humans
  • Intubation, Intratracheal / instrumentation*
  • Male
  • Middle Aged
  • Prospective Studies
  • Thoracic Surgical Procedures