Development of a primary thoron activity standard for the calibration of thoron measurement instruments

Radiat Prot Dosimetry. 2015 Nov;167(1-3):70-4. doi: 10.1093/rpd/ncv221. Epub 2015 Apr 24.

Abstract

The LNHB and IRSN are working on a reference atmosphere for thoron ((220)Rn) instrument calibration. The LNHB, as the national metrology institute for activity measurement in France, has to create a new thoron reference standard in order to estimate with accuracy the thoron concentration of a reference atmosphere. The measurement system presented in this paper is based on a reference volume using an alpha detector, which is able to measure thoron and its decay products to define the thoron concentration of a thoron reference atmosphere. This paper presents the first results with this new system using a well-known radon ((222)Rn) atmosphere and a thoron ((220)Rn) atmosphere.