An Irrigation Suction System Using a Pressurized Fluid High-Flow Pump System in Comparison with a Standard Cuff Pressure System: Technical Note

J Neurol Surg A Cent Eur Neurosurg. 2016 May;77(3):264-8. doi: 10.1055/s-0034-1376188. Epub 2015 Apr 27.

Abstract

Objective: In neurosurgical operations, such as clipping of ruptured cerebral aneurysms, irrigation of the operative field is a basic and extremely important surgical technique. Various irrigation-suction (IS) systems have been marketed for irrigation of operative fields in neurosurgery. However, with these devices, we often must increase the cuff pressure to maintain fluid power and change the fluid bag relatively frequently. We report the usefulness of pressurized irrigation fluid with the Niagara high-flow irrigation system (NI).

Method: Using both the NI system and an IS, we altered pressure and duration to investigate differences in flow rate.

Results: There was no change in flow rate 80 minutes after pressurization of the NI system, and flow rates decreased slightly. Additionally, flow rates using NI were sensitive to changes in pressure.

Conclusion: This system has reduced the workload of assisting nurses and surgeons during surgery. Thus we consider the technique to be very useful for neurosurgical operations that require delicate procedures.

MeSH terms

  • Equipment Design*
  • Humans
  • Neurosurgical Procedures / instrumentation*
  • Suction / instrumentation*
  • Suction / methods
  • Therapeutic Irrigation / instrumentation*