Reproducibility of DCE-MRI time-intensity curve-shape analysis in patients with knee arthritis: A comparison with qualitative and pharmacokinetic analyses

J Magn Reson Imaging. 2015 Dec;42(6):1497-506. doi: 10.1002/jmri.24933. Epub 2015 Apr 29.

Abstract

Purpose: To compare the between-session reproducibility of dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) combined with time-intensity curve (TIC)-shape analysis in arthritis patients, within one scanner and between two different scanners, and to compare this method with qualitative analysis and pharmacokinetic modeling (PKM).

Materials and methods: Fifteen knee joint arthritis patients were included and scanned twice on a closed-bore 1.5T scanner (n = 9, group 1), or on a closed-bore 1.5T and on an open-bore 1.0T scanner (n = 6, group 2). DCE-MRI data were postprocessed using in-house developed software ("Dynamo"). Disease activity was assessed.

Results: Disease activity was comparable between the two visits. In group 1 qualitative analysis showed the highest reproducibility with intraclass correlation coefficients (ICCs) between 0.78 and 0.98 and root mean square-coefficients of variation (RMS-CoV) of 8.0%-14.9%. TIC-shape analysis showed a slightly lower reproducibility with similar ICCs (0.78-0.97) but higher RMS-CoV (18.3%-42.9%). The PKM analysis showed the lowest reproducibility with ICCs between 0.39 and 0.64 (RMS-CoV 21.5%-51.9%). In group 2 TIC-shape analysis of the two most important TIC-shape types showed the highest reproducibility with ICCs of 0.78 and 0.71 (RMS-CoV 29.8% and 59.4%) and outperformed the reproducibility of the most important qualitative parameter (ICC 0.31, RMS-CoV 45.1%) and the within-scanner reproducibility of PKM analysis.

Conclusion: TIC-shape analysis is a robust postprocessing method within one scanner, almost as reproducible as the qualitative analysis. Between scanners, the reproducibility of the most important TIC-shapes outperform that of the most important qualitative parameter and the within-scanner reproducibility of PKM analysis.

Keywords: DCE-MRI; Ktrans; TIC-shape; arthritis; knee.

Publication types

  • Comparative Study
  • Controlled Clinical Trial

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Algorithms*
  • Computer Simulation
  • Contrast Media / pharmacokinetics
  • Female
  • Gadolinium DTPA / pharmacokinetics*
  • Humans
  • Image Interpretation, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / methods
  • Male
  • Metabolic Clearance Rate
  • Middle Aged
  • Models, Biological*
  • Osteoarthritis, Knee / metabolism*
  • Osteoarthritis, Knee / pathology*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Contrast Media
  • Gadolinium DTPA