Development of an Ex Vivo, Beating Heart Model for CT Myocardial Perfusion

Biomed Res Int. 2015:2015:412716. doi: 10.1155/2015/412716. Epub 2015 Jun 21.

Abstract

Objective: To test the feasibility of a CT-compatible, ex vivo, perfused porcine heart model for myocardial perfusion CT imaging.

Methods: One porcine heart was perfused according to Langendorff. Dynamic perfusion scanning was performed with a second-generation dual source CT scanner. Circulatory parameters like blood flow, aortic pressure, and heart rate were monitored throughout the experiment. Stenosis was induced in the circumflex artery, controlled by a fractional flow reserve (FFR) pressure wire. CT-derived myocardial perfusion parameters were analysed at FFR of 1 to 0.10/0.0.

Results: CT images did not show major artefacts due to interference of the model setup. The pacemaker-induced heart rhythm was generally stable at 70 beats per minute. During most of the experiment, blood flow was 0.9-1.0 L/min, and arterial pressure varied between 80 and 95 mm/Hg. Blood flow decreased and arterial pressure increased by approximately 10% after inducing a stenosis with FFR ≤ 0.50. Dynamic perfusion scanning was possible across the range of stenosis grades. Perfusion parameters of circumflex-perfused myocardial segments were affected at increasing stenosis grades.

Conclusion: An adapted Langendorff porcine heart model is feasible in a CT environment. This model provides control over physiological parameters and may allow in-depth validation of quantitative CT perfusion techniques.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Coronary Angiography / methods
  • Coronary Stenosis / diagnostic imaging*
  • Disease Models, Animal*
  • Humans
  • Isolated Heart Preparation / methods*
  • Myocardial Perfusion Imaging / methods*
  • Organ Culture Techniques / methods*
  • Perfusion / methods
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Swine
  • Tomography, X-Ray Computed / methods*