Purpose: To investigate image quality, radiation dose, and diagnostic performance of prospectively ECG-triggered high-pitch coronary CT angiography (CCTA) at 70 kVp compared to invasive coronary angiography (ICA) as reference standard.
Materials and methods: Forty-three patients underwent prospectively ECG-triggered high-pitch CCTA at 70 kVp using 30 cc (11 g iodine) contrast medium and ICA. Subjective and objective image quality was evaluated for each CCTA study. CCTA performance for diagnosing ≥50% stenosis was assessed. Results were stratified according to heart rate (HR), body mass index (BMI), Agatston score, and image quality.
Results: At CCTA, 94.3% (500/530) of coronary segments were of diagnostic quality. Using ICA as reference standard, sensitivity and accuracy were 100% and 93.0% on a per-patient basis. Per-vessel and per-segment performances were 92.2% and 89.5%; 79.5% and 88.3%, respectively. No differences were found in diagnostic accuracy between different HR, BMI, and calcification subgroups (all P > 0.05) on a per-patient basis. However, low image quality reduced diagnostic accuracy on a per-patient, per-vessel and per-segment basis (all P < 0.05). The mean effective radiation dose was 0.2 ± 0.0 mSv.
Conclusion: Our presented protocol results in an effective radiation dose of 0.2 mSv and high diagnostic accuracy for stenosis detection in a selected, non-obese population.
Key points: Prospectively ECG-triggered high-pitch CCTA at 70 kVp is feasible. This protocol has a high diagnostic accuracy for stenosis detection. The mean effective radiation dose was 0.2 ± 0.0 mSv. Only 30 cc of contrast material is used in this protocol. Low image quality reduced diagnostic accuracy of CCTA.
Keywords: Coronary CT angiography; High-pitch acquisition; Low contrast medium; Low kilovoltage; Prospective ECG gating.