Drug-Coated Balloons Versus Second-Generation Drug-Eluting Stents for the Management of Recurrent Multimetal-Layered In-Stent Restenosis

JACC Cardiovasc Interv. 2015 Oct;8(12):1586-94. doi: 10.1016/j.jcin.2015.04.032. Epub 2015 Sep 17.

Abstract

Objectives: This study aimed to investigate the clinical outcomes of patients presenting with recurrent drug-eluting stent (DES) in-stent restenosis (ISR) treated with a second-generation DES or with a drug-coated balloon (DCB).

Background: To date, there are no reports of DCB treatment and limited data with regard to the efficacy of further DES implantation for recurrent ISR.

Methods: Between January 2008 and December 2013, 171 lesions were assessed for eligibility (82 lesions in the second-generation DES group and 89 lesions in the DCB group).

Results: Acute gain was greater in the second-generation DES group (second-generation DES, 2.09 ± 0.53 mm vs. DCBs, 1.60 ± 0.62 mm, p < 0.001). The rates of major adverse cardiac events were comparable (at 1 year, DES 14.0% vs. DCBs 12.3%; at 2 years, DES 28.8% vs. DCBs 43.5%, p = 0.21). Major adverse cardiac event rates were mainly driven by target lesion revascularization (at 1 year, DES 12.5% vs. DCBs 10.9%; at 2 years, DES 27.7% vs. DCBs 38.3%; p = 0.40). Definite scaffold thrombosis occurred in 2 patients (1 patient in each group). Multivariable analysis revealed ISR recurrence within 1 year (hazard ratio: 2.43, 95% confidence interval: 1.14 to 5.18, p = 0.02) and lesion length (per 10-mm increase) (hazard ratio: 1.15, 95% confidence interval: 1.00 to 1.32, p = 0.049) to be independent predictors of TLR.

Conclusions: The results after both treatments were equivalent. ISR recurrence within 1 year of the first reintervention and lesion length were independent predictors of future target lesion revascularization. Larger studies are required to confirm the late (>1 year) differences with regard to clinical outcomes.

Keywords: drug-coated balloon; drug-eluting stent; in-stent restenosis.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Angioplasty, Balloon, Coronary / mortality
  • Cardiac Catheters*
  • Chi-Square Distribution
  • Coated Materials, Biocompatible*
  • Coronary Angiography
  • Coronary Restenosis / diagnosis
  • Coronary Restenosis / mortality
  • Coronary Restenosis / therapy*
  • Drug-Eluting Stents*
  • Female
  • Humans
  • Italy
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Multivariate Analysis
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / mortality
  • Proportional Hazards Models
  • Prosthesis Design
  • Risk Factors
  • Time Factors
  • Treatment Outcome

Substances

  • Coated Materials, Biocompatible