Portable and compact instruments for separating and detecting organic compounds are needed in the field for environmental studies. This is especially the case for pollution studies as in-field detection of organic compounds helps identify sources of pollution. Here we describe the development of a compact GC and simple pre-concentrator coupled to a MS detector. This simple system can easily be incorporated into portable instrumentation. Combining the pre-concentrator and compact column has the advantage of decoupling separation from manual injection and enhances separation of environmentally relevant polar organic compounds, such as levoglucosan. A detection limit of 2.2 ng was obtained for levoglucosan. This simple design has the potential to expand the use of gas chromatography as a routine in-field separation technique.
Keywords: Compact GC; Field sampling; Levoglucosan; Pre-concentration.
Copyright © 2015 Elsevier B.V. All rights reserved.