We demonstrate the feasibility of using optical coherence tomography (OCT) to image and detect 2.8 μm diameter microparticles (stationary and moving) on a highly-reflective gold surface both in clear media and under skin in vitro. The OCT intensity signal can clearly report the microparticle count, and the OCT response to the number of microparticles shows a good linearity. The detect ability of the intensity change (2.9% ± 0.5%) caused by an individual microparticle shows the high sensitivity of monitoring multiple particles using OCT. An optical sensing method based on this feasibility study is described for continuously measuring blood sugar levels in the subcutaneous tissue, and a molecular recognition unit is designed using competitive binding to modulate the number of bound microparticles as a function of glucose concentration. With further development, an ultra-small, implantable sensor might provide high specificity and sensitivity for long-term continuous monitoring of blood glucose concentration.
Keywords: Biomedical monitoring; magnetic microparticles; optical coherence tomography; retroreflection; skin.