The study of the influence of the lipemia and icterus was performed experimentally for twenty-four biochemistry parameters on the Roche Cobas 6000 CE analyzer. Overloads in Intralipid(®) or ditaurate of bilirubin were performed on several plasma pools. The limit of 10% was chosen to define interference on the measurement. The parameters studied were classified into different categories depending on their measurement is affected or not. Knowledge of these data allows the biologist to adapt its reporting procedures in the case of lactescent and/or icteric samples.
Keywords: bilirubin; icterus; interference; lactescence; lipemia.