In spite of the increasing applications of luminescence measurements, no instrumentation is yet commercially available that permits sensitive luminescence measurements to be performed simultaneously on a large number of samples. Based on the principle of single-photon imaging, we describe here the performance of a computer-aided image luminometer with a capacity for two microtiter plates. Applications to the study of chemiluminescence by a B lymphocyte cell line, and to an IgE luminescence immunoassay are used to exemplify the capabilities of the system, which we have termed 'chemiluminescence multiwell analyser'.