Standardisation and precise determination of the half-life of (44)Sc

Appl Radiat Isot. 2016 Mar:109:314-318. doi: 10.1016/j.apradiso.2015.12.007. Epub 2015 Dec 5.

Abstract

The half-life of the positron-emitter (44)Sc has been determined by following the decay rate with two measurement systems; an Ionisation Chamber and a HPGe detector. The combination of seven results gives a value of T1/2=4.042 (25)h, about 2% higher than the recommended value of T1/2=3.97 (4)h (Browne, 2011) and with a lower uncertainty. This radionuclide has also been standardised by coincidence counting, and liquid scintillation counting techniques. A (44)Ti/(44)Sc generator developed at CIEMAT was used to obtain the (44)Sc solutions used in all measurements.

Keywords: Half-life; Positron emitters; Sc-44; Standardisation.