Quantification of velocity reduction after flow diverter placement in intracranial aneurysm: An ex vivo study with 3D printed replicas

Annu Int Conf IEEE Eng Med Biol Soc. 2015:2015:7300-3. doi: 10.1109/EMBC.2015.7320077.

Abstract

Phase contrast MRI (pcMRI) was used to measure flow before and after placement of a flow diverter (n = 3). Decreases from 18% to 31% in flow velocity were seen in the inflow jet of the aneurysms. Flow patterns were also compared. It was observed that the gross aneurysmal flow patterns were maintained after flow diverter placement despite decreased fluid velocities. All measurements were carried out in 3D printed aneurysm replicas.

MeSH terms

  • Blood Flow Velocity
  • Humans
  • Intracranial Aneurysm / physiopathology*
  • Intracranial Aneurysm / therapy*
  • Magnetic Resonance Imaging
  • Printing, Three-Dimensional*
  • Replica Techniques*