Purpose: To characterize the q-space truncation and sampling on the spin-displacement probability density function (PDF) in diffusion spectrum imaging (DSI).
Methods: DSI data were acquired using the MGH-USC connectome scanner (Gmax = 300 mT/m) with bmax = 30,000 s/mm2 , 17 × 17 × 17, 15 × 15 × 15 and 11 × 11 × 11 grids in ex vivo human brains and bmax = 10,000 s/mm2 , 11 × 11 × 11 grid in vivo. An additional in vivo scan using bmax =7,000 s/mm2 , 11 × 11 × 11 grid was performed with a derated gradient strength of 40 mT/m. PDFs and orientation distribution functions (ODFs) were reconstructed with different q-space filtering and PDF integration lengths, and from down-sampled data by factors of two and three.
Results: Both ex vivo and in vivo data showed Gibbs ringing in PDFs, which becomes the main source of artifact in the subsequently reconstructed ODFs. For down-sampled data, PDFs interfere with the first replicas or their ringing, leading to obscured orientations in ODFs.
Conclusion: The minimum required q-space sampling density corresponds to a field-of-view approximately equal to twice the mean displacement distance (MDD) of the tissue. The 11 × 11 × 11 grid is suitable for both ex vivo and in vivo DSI experiments. To minimize the effects of Gibbs ringing, ODFs should be reconstructed from unfiltered q-space data with the integration length over the PDF constrained to around the MDD. Magn Reson Med 76:1750-1763, 2016. © 2016 International Society for Magnetic Resonance in Medicine.
Keywords: Gibbs ringing artifacts; aliasing artifacts; diffusion spectrum imaging; ex vivo imaging; high b-value; q-space truncation and sampling.