Feasibility of Non-contrast-enhanced MR Angiography Using the Time-SLIP Technique for the Assessment of Pulmonary Arteriovenous Malformation

Magn Reson Med Sci. 2016 Jul 11;15(3):253-65. doi: 10.2463/mrms.mp.2015-0069. Epub 2016 Feb 3.

Abstract

Purpose: The purpose of this study was to evaluate the diagnostic performance of non-contrast-enhanced magnetic resonance angiography with time-spatial labeling inversion pulse (time-SLIP MRA) in the assessment of pulmonary arteriovenous malformation (PAVM).

Methods: Eleven consecutive patients with 38 documented PAVMs underwent time-SLIP MRA with a 3-tesla unit. Eight patients with 25 lesions were examined twice, once before and once after embolotherapy. The lesions were divided into two groups-initial diagnosis (n = 35) and follow-up (n = 28)-corresponding to untreated and treated lesions, respectively, and were evaluated separately. To evaluate the initial diagnosis group, two reviewers assessed image quality for visualization of PAVMs by using a qualitative 4-point scale (1 = not assessable to 4 = excellent). The location and classification of PAVMs were also evaluated. The results were compared with those from digital subtraction angiography. For evaluation of the follow-up group, the reviewers assessed the status of treated PAVMs. Reperfusion and occlusion were defined respectively as visualization or disappearance of the aneurysmal sac. The diagnostic accuracy of time-SLIP MRA was assessed and compared with standard reference images. Interobserver agreement was evaluated with the κ statistic.

Results: In the initial diagnosis group, time-SLIP MRA correctly determined the PAVMs in all but one patient with one lesion who had image degradation due to irregular breath. Image quality was considered excellent (median = 4) and the κ coefficient was 0.85. Additionally, both readers could correctly localize and classify the PAVMs on time-SLIP MRA images with both κ coefficient of 1.00. In the follow-up group, the sensitivity and specificity of time-SLIP MRA for reperfusion of PAVMs were both 100%, and the κ coefficient was 1.00.

Conclusion: Time-SLIP MRA is technically and clinically feasible and represents a promising technique for noninvasive pre- and post-treatment assessment of PAVMs.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Angiography, Digital Subtraction / methods*
  • Arteriovenous Fistula / diagnostic imaging*
  • Contrast Media
  • Female
  • Humans
  • Magnetic Resonance Angiography / methods*
  • Male
  • Middle Aged
  • Prospective Studies
  • Pulmonary Artery / abnormalities*
  • Pulmonary Artery / diagnostic imaging
  • Pulmonary Veins / abnormalities*
  • Pulmonary Veins / diagnostic imaging
  • Sensitivity and Specificity
  • Young Adult

Substances

  • Contrast Media

Supplementary concepts

  • Pulmonary Arteriovenous Fistulas