StatNet Electroencephalogram: A Fast and Reliable Option to Diagnose Nonconvulsive Status Epilepticus in Emergency Setting

Can J Neurol Sci. 2016 Mar;43(2):254-60. doi: 10.1017/cjn.2015.391. Epub 2016 Feb 11.

Abstract

Background: The StatNet electrode set is a system that can be applied by a non-electroencephalogram (EEG) technologist after minimal training. The primary objectives of this study are to assess the quality and reliability of the StatNet recordings in comparison to the conventional EEG.

Methods: Over 10 months, 19 patients with suspected nonconvulsive status epilepticus were included from university hospital emergency settings. Each patient received a StatNet EEG by a trained epilepsy fellow and a conventional EEG by registered technologists. We compared the studies in a blinded fashion, for the timeframe from EEG order to the setup time, start of acquisition, amount of artifact, and detection of abnormalities. The nonparametric Mann-Whitney two-sample t test was used for comparisons. The kappa score was used to assess reliability.

Results: Mean age of patients was 61±16.3 (25-93) years. The inter-observer agreement for detection of abnormal findings was 0.83 for StatNet and 0.75 for conventional EEG. Nonconvulsive status epilepticus was detected in 10% (2/19) in both studies. The delay from the time of EEG requisition to acquisition was shorter in the StatNet (22.4±2.5 minutes) than the conventional EEG (217.7±44.6 minutes; p<0.0001). The setup time was also shorter in the StatNet (9.9±0.8 minutes) compared with the conventional EEG (17.8±0.8 minutes; p<0.0001). There was no difference in the percentage of artifact duration between the two studies (p=0.89).

Conclusion: This study demonstrates that StatNet EEG is a practical and reliable tool in the emergency setting, which reduces the delay of testing compared with conventional EEG, without significant compromise of study quality.

Keywords: disposable EEG electrodes; emergency department; emergent EEG; intensive care unit; nonconvulsive status epilepticus; stat EEG.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Electroencephalography / methods*
  • Emergency Service, Hospital
  • Female
  • Humans
  • Male
  • Middle Aged
  • Reproducibility of Results
  • Signal Processing, Computer-Assisted
  • Status Epilepticus / diagnosis*