Durability after aortic valve replacement with the Mitroflow versus the Perimount pericardial bioprosthesis: a single-centre experience in 2393 patients

Eur J Cardiothorac Surg. 2016 Jun;49(6):1705-10. doi: 10.1093/ejcts/ezv432. Epub 2016 Mar 16.

Abstract

Objectives: This study compares the durability and risk of reoperation in patients undergoing aortic valve replacement (AVR) with either a Mitroflow or a Carpentier-Edwards (CE) pericardial bioprosthesis. Since AVR with bioprosthetic valves has increased progressively in recent years as compared to mechanical valves, especially in patients aged 60-70 years, there has been renewed interest in the long-term durability of current pericardial bioprostheses.

Methods: We compared 440 AVR with Mitroflow valves with 1953 AVR with CE pericardial valves implanted from 1999 to 2014 with regard to reoperation, reoperation for structural valve deterioration (SVD) and all-cause mortality.

Results: Ten-year freedom from explant of any cause was higher for CE Perimount (98 ± 0.7%) than for Mitroflow (95 ± 1.4%, P < 0.01). Reasons for explant for CE Perimount were SVD (n = 2), endocarditis (n = 8) and paraprosthetic leak (n = 10). The reasons for explant for Mitroflow were SVD (n = 11), endocarditis (n = 3) SVD and pericarditis (n = 1) and paraprosthetic leak (n = 2). Ten-year freedom from explant due to SVD was higher for CE Perimount (100%) than for Mitroflow (96%) (P < 0.01). In small aortic annuli (bioprosthesis size 19-21 mm), freedom from SVD at 10 years for CE Perimount and Mitroflow was 100 versus 96%, respectively. By multivariate analysis, it was found that bioprosthesis size was not a risk factor for SVD. The choice of valve type could not be demonstrated to influence long-term survival.

Conclusions: The Mitroflow pericardial bioprosthesis provides less than optimal mid- and long-term durability compared with the CE Perimount pericardial valve, especially for small aortic diameter implants (19 and 21 mm). This study hereby confirms the existence of a real risk of valvular deterioration of the Mitroflow valve that might compromise the prognosis of the patients.

Keywords: Aortic valve replacement; Structural valve deterioration.

Publication types

  • Observational Study

MeSH terms

  • Aged
  • Aortic Valve / surgery*
  • Bioprosthesis*
  • Endocarditis / surgery
  • Female
  • Heart Valve Diseases / surgery*
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / methods
  • Heart Valve Prosthesis Implantation / statistics & numerical data
  • Heart Valve Prosthesis*
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Pericardium / transplantation
  • Prosthesis Design
  • Prosthesis Failure
  • Prosthesis-Related Infections / surgery
  • Reoperation / methods
  • Reoperation / statistics & numerical data