The effect of time in use on the display performance of the iPad

Br J Radiol. 2016 Jul;89(1063):20150657. doi: 10.1259/bjr.20150657. Epub 2016 May 16.

Abstract

Objective: The aim of this study was to evaluate changes to the luminance, luminance uniformity and conformance to the digital imaging and communication in medicine greyscale standard display function (GSDF) as a function of time in use for the iPad.

Methods: Luminance measurements of the American Association of Physicists in Medicine (AAPM) Group 18 task group (TG18) luminance uniformity and luminance test patterns (TG18-UNL and TG18-LN8) were performed using a calibrated near-range luminance meter. Nine sets of measurements were taken, where the time in use of the iPad ranged from 0 to 2500 h.

Results: The maximum luminance (Lmax) of the display decreased (367-338 cdm(-2)) as a function of time. The minimum luminance remained constant. The maximum non-uniformity coefficient was 11%. Luminance uniformity decreased slightly as a function of time in use. The conformance of the iPad deviated from the GSDF curve at commencement of use. Deviation did not increase as a function of time in use.

Conclusion: This study has demonstrated that the iPad display exhibits luminance degradation typical of liquid crystal displays. The Lmax of the iPad fell below the American College of Radiology-AAPM-Society of Imaging Informatics in Medicine recommendations for primary displays (>350 cdm(-2)) at approximately 1000 h in use. The Lmax recommendation for secondary displays (>250 cdm(-2)) was exceeded during the entire study. The maximum non-uniformity coefficient did not exceed the recommendations for either primary or secondary displays. The deviation from the GSDF exceeded the recommendations of the TG18 for use as either a primary or secondary display.

Advances in knowledge: The brightness, uniformity and contrast response are reasonably stable over the useful lifetime of the device; however, the device fails to meet the contrast response standard for either a primary or secondary display.

MeSH terms

  • Data Display / standards*
  • Humans
  • Liquid Crystals
  • Microcomputers*
  • Time