Objectives: To prospectively evaluate the feasibility, safety and accuracy of MR-guided percutaneous biopsy of solitary pulmonary lesions using a 1.0-T open MR scanner with respiratory gating.
Methods: Sixty-five patients with 65 solitary pulmonary lesions underwent MR-guided percutaneous coaxial cutting needle biopsy using a 1.0-T open MR scanner with respiratory gating. Lesions were divided into two groups according to maximum lesion diameters: ≤2.0 cm (n = 31) and >2.0 cm (n = 34). The final diagnosis was established in surgery and subsequent histology. Diagnostic accuracy, sensitivity and specificity were compared between the groups using Fisher's exact test.
Results: Accuracy, sensitivity and specificity of MRI-guided percutaneous pulmonary biopsy in diagnosing malignancy were 96.9 %, 96.4 % and 100 %, respectively. Accuracy, sensitivity and specificity were 96.8 %, 96.3 % and 100 % for lesions 2.0 cm or smaller and 97.1 %, 96.4 % and 100 %, respectively, for lesions larger than 2.0 cm. There was no significant difference between the two groups (P > 0.05). Biopsy-induced complications encountered were pneumothorax in 12.3 % (8/65) and haemoptysis in 4.6 % (3/65). There were no serious complications.
Conclusions: MRI-guided percutaneous biopsy using a 1.0-T open MR scanner with respiratory gating is an accurate and safe diagnostic technique in evaluation of pulmonary lesions.
Key points: • MRI-guided percutaneous lung biopsy using a 1.0-T open MR scanner is feasibility. • 96.9 % differentiation accuracy of malignant and benign lung lesions is possible. • No serious complications occurred in MRI-guided lung biopsy.
Keywords: Guidance; Interventional radiology; Lung biopsy; Magnetic resonance imaging; Technology.