In this study, we present a case of falsely elevated oestradiol (E2) concentration, determined by two immunoassays, in a breast cancer patient receiving exemestane therapy. The positive bias of immunochemical measurements was revealed using liquid chromatography tandem mass spectrometry which showed undetectable E2 concentration. The discrepancy is expected to be a consequence of the structural resemblance of E2 and exemestane sharing the same steroidal backbone. Inaccurate laboratory findings in therapy monitoring, as in this case, may lead to unnecessary changes of therapy.
Keywords: Immunoassay; mass spectrometry; steroid hormones.