Safety and efficacy of applying a low-dose radiation fluoroscopy protocol in device implantations

Europace. 2017 Aug 1;19(8):1364-1368. doi: 10.1093/europace/euw189.

Abstract

Aims: For cardiac implantable electronic device (CIED) implantations, visualization of lead placement is necessary and fluoroscopy remains by far the most commonly used technique. With simple changes in the X-ray system settings, total radiation dose can be reduced significantly. The purpose of this study was to assess the safety and efficacy of various CIED implantations performed after implementation of a new dose reduction protocol (DRP).

Methods and results: We conducted a retrospective chart review of 584 patients undergoing CIED implantation or revision in our hospital. Of these patients, 280 (48%) underwent the implantation prior to and 304 (52%) after the DRP introduction. The DRP included various changes for optimized image processing and exposure system settings to enable dose reduction, as well as a reduced frame rates (4 FPS for fluoroscopy and 7.5 FPS for cinematographic images). Of the 584 patients, 53 (9.1%) had a one-chamber pacemaker, 232 (39.7%) a two-chamber pacemaker, 133 (22.8%) a one-chamber ICD, 35 (6.0%) a two-chamber ICD, 82 (14.0%) a CRT (de novo) implantation, and 49 (8.3%) had an upgrade to a CRT device. DRP was associated with a 64% reduction of the dose-area product (1372 ± 2659 vs. 3792 ± 5025 cGcm2, P < 0.001), while fluoroscopy duration (13 ± 15 vs. 13 ± 15 min) and procedural duration (93 ± 52 vs. 92 ± 52 min.) did not significantly increase. Complication rates did not differ significantly between the two groups.

Conclusion: The DRP proved to effectively reduce radiation dose for all types of CIED implantations. Fluoroscopy time, total procedure time, and the number of complications did not increase after introducing the DRP.

Keywords: CIED implantations; Complications; Dose reduction; Fluoroscopy.

Publication types

  • Comparative Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Cardiac Resynchronization Therapy Devices
  • Defibrillators, Implantable*
  • Female
  • Fluoroscopy
  • Germany
  • Humans
  • Male
  • Middle Aged
  • Pacemaker, Artificial*
  • Patient Safety
  • Program Evaluation
  • Prosthesis Design
  • Prosthesis Implantation / adverse effects
  • Prosthesis Implantation / instrumentation*
  • Prosthesis Implantation / methods*
  • Protective Factors
  • Radiation Dosage*
  • Radiation Exposure / adverse effects
  • Radiation Exposure / prevention & control*
  • Radiography, Interventional* / adverse effects
  • Retrospective Studies
  • Risk Assessment
  • Risk Factors
  • Young Adult