Measurement of helium-like and hydrogen-like argon spectra using double-crystal X-ray spectrometers on EAST

Rev Sci Instrum. 2016 Nov;87(11):11E326. doi: 10.1063/1.4960504.

Abstract

A two-crystal assembly was deployed on the tangential X-ray crystal spectrometer to measure both helium-like and hydrogen-like spectra on EAST. High-quality helium-like and hydrogen-like spectra were observed simultaneously for the first time on one detector for a wide range of plasma parameters. Profiles of line-integrated core ion temperatures inferred from two spectra were consistent. Since tungsten was adopted as the upper divertor material, one tungsten line (W XLIV at 4.017 Å) on the short-wavelength side of the Lyman-α line (Lα1) was identified for typical USN discharges, which was diffracted by a He-like crystal (2d = 4.913 Å). Another possible Fe XXV line (1.85 Å) was observed to be located on the long-wavelength side of resonance line (w), which was diffracted from a H-like crystal (2d = 4.5622 Å) on the second order. Be-like argon lines were also observable that fill the detector space between the He-like and H-like spectra.