MD-Logic overnight type 1 diabetes control in home settings: A multicentre, multinational, single blind randomized trial

Diabetes Obes Metab. 2017 Apr;19(4):553-561. doi: 10.1111/dom.12852. Epub 2017 Jan 19.

Abstract

Aims: To evaluate the safety, efficacy and need for remote monitoring of the MD-Logic closed-loop system during short-term overnight use at home.

Methods: Seventy-five patients (38 male; aged 10-54 years; average A1c, 7.8% ± 0.7%, 61.8 ± 7.2 mmol/mol) were enrolled from 3 clinical sites. Patients were randomly assigned to participate in 2 overnight crossover periods, each including 4 consecutive nights, 1 under closed-loop control and 1 under sensor-augmented pump (SAP) therapy in the patient's home. Both study arms were supervised using a remote-monitoring system in a blinded manner. Primary endpoints were time spent with glucose levels below 70 mg/dL and percentage of nights in which mean overnight glucose levels were within 90 to 140 mg/dL.

Results: The median [interquartile range] percentage of time spent in hypoglycaemia was significantly lower on nights when MD-Logic was used, compared to SAP therapy (2.07 [0, 4.78] and 2.6 [0, 10.34], respectively; P = .004) and the percentage of individual nights with a mean overnight glucose level in target was significantly greater (75 [42, 75] and 50 [25,75], respectively; P = .008). The time spent in target range was increased by a median of 28% (P = .001), with the same amount of insulin (10.69 [7.28, 13.94] and 10.41[6.9, 14.07], respectively; P = .087). The remote monitoring triggered calls for hypoglycaemia at twice the rate during SAP therapy compared to closed-loop control (62 and 29, respectively; P = .002).

Conclusions: The MD-Logic system demonstrated a safe and efficient profile during overnight use by children, adolescents and adults with type 1 diabetes and, therefore, provides an effective means of mitigating the risk of nocturnal hypoglycaemia.

Keywords: MD-Logic; closed-loop; nocturnal hypoglycaemia; remote-monitoring; type 1.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Blood Glucose / analysis
  • Blood Glucose Self-Monitoring / instrumentation*
  • Child
  • Cross-Over Studies
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / complications
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Drug Chronotherapy
  • Female
  • Humans
  • Hypoglycemia / blood
  • Hypoglycemia / drug therapy
  • Hypoglycemia / etiology
  • Hypoglycemic Agents / administration & dosage*
  • Insulin / administration & dosage*
  • Insulin Infusion Systems*
  • Male
  • Middle Aged
  • Single-Blind Method
  • Telemetry / methods
  • Time Factors
  • Treatment Outcome
  • Young Adult

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin