Comparative assessment of "plaque/media" change on three modalities of IVUS immediately after implantation of either everolimus-eluting bioresorbable vascular scaffold or everolimus-eluting metallic stent in Absorb II study

Int J Cardiovasc Imaging. 2017 Apr;33(4):441-449. doi: 10.1007/s10554-016-1033-7. Epub 2016 Dec 23.

Abstract

The purpose of the study to assess the comparability of immediate changes in plaque/media volume (PV) on three modalities of intravascular ultrasound (IVUS) after implantation of either bioresorbable vascular scaffold (BVS) or everolimus-eluting metallic stent (EES) in Absorb II Study. The two devices have different device volume and ultrasound backscattering that may interfere with the "plaque/media" assessed by three modalities on IVUS: grayscale, backscattering of radiofrequency and brightness function. In a multicenter randomized controlled trial, 501 patients with stable or unstable angina underwent documentary IVUS pre- and post- implantation. The change in plaque/media volume (PV) was categorized into three groups according to the relative PV change in device segment: PV "increased" >+5% (PVI), PV unchanged ±5% (PVU), and PV decreased <-5% (PVD). The change in PV was re-evaluated three times: after subtraction of theoretical device volume, after analysis of echogenicity based on brightness function. In 449 patients, 483 lesions were analyzed pre- and post-implantation. "PVI" was more frequently observed in BVS (53.8%) than EES group (39.4%), p = 0.006. After subtraction of the theoretical device volume, the frequency of "PVI" decreased in both BVS (36.2%) and EES (32.1%) groups and became comparable (p = 0.581). In addition, the percentage of "PVI" was further reduced in both device groups after correction for either radiofrequency backscattering (BVS 34.4% vs. EES 22.6%) or echogenicity (BVS 25.2% vs. EES 9.7%). PV change in device segment was differently affected by BVS and EES devices implantation due to their differences in device volume and ultrasound backscattering. It implies that the lumen volume was also artifactually affected by the type of device implanted. Comparative IVUS assessment of lumen and plaque/media volume changes following implantation of BVS and EES requires specific methodological adjustment.

Keywords: Absorb II; Bioresorbable vascular scaffold; IVUS; Plaque/media volume.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Absorbable Implants*
  • Aged
  • Cardiovascular Agents / administration & dosage*
  • Cardiovascular Agents / adverse effects
  • Carotid Artery Diseases / diagnostic imaging*
  • Carotid Artery Diseases / pathology
  • Carotid Artery Diseases / therapy*
  • Coronary Vessels / diagnostic imaging*
  • Coronary Vessels / pathology
  • Databases, Factual
  • Drug-Eluting Stents*
  • Everolimus / administration & dosage*
  • Everolimus / adverse effects
  • Female
  • Humans
  • Image Interpretation, Computer-Assisted
  • Male
  • Metals*
  • Middle Aged
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Plaque, Atherosclerotic*
  • Predictive Value of Tests
  • Prospective Studies
  • Prosthesis Design
  • Single-Blind Method
  • Time Factors
  • Treatment Outcome
  • Ultrasonography, Interventional*

Substances

  • Cardiovascular Agents
  • Metals
  • Everolimus