Stentless vs Stented Aortic Valve Bioprostheses in the Small Aortic Root

Semin Thorac Cardiovasc Surg. 2016;28(2):390-397. doi: 10.1053/j.semtcvs.2016.02.012. Epub 2016 Mar 2.

Abstract

In patients with a small aortic root undergoing aortic valve replacement (AVR), the Freedom SOLO bioprosthesis may be the ideal prosthesis because of its stentless design and supra-annular implantation. This study investigated if the stentless Freedom SOLO has an advantage when compared with a stented bioprosthesis in patients with a small aortic root. From April 2005-July 2014, 269 consecutive patients underwent AVR with either a Freedom SOLO (n = 76) or Mitroflow (n = 193) bioprosthesis size 19mm or 21mm, respectively. This retrospective comparison study presents clinical and echocardiographic follow-up data. In results, operative outcome and survival were similar. At 7 years, cumulative incidence of aortic valve reoperation and structural valve deterioration favor the Freedom SOLO (0% vs 7.1%, P = 0.03 and 0% vs 4.5%, P = 0.08, respectively). Additionally, the postoperative peak and mean valvular gradients favor the Freedom SOLO (21 ± 9mmHg vs 32 ± 12mmHg and 12 ± 5mmHg vs 19 ± 8mmHg, both P = <0.001, respectively). During mid-term follow-up this hemodynamic advantage continued in favor of the Freedom SOLO. Also prosthesis-patient mismatch occurred less frequently in the Freedom SOLO (28% vs 52%, P = 0.001). There were no differences in prosthetic valve endocarditis, thromboembolic, or bleeding events. In conclusion, the stentless Freedom SOLO has several significant advantages for AVR in patients with a small aortic root in comparison with a stented Mitroflow bioprosthesis. The Freedom SOLO shows superior hemodynamic performance with significantly lower valvular gradients that remained stable during mid-term follow-up. Additionally, significantly fewer prosthesis-patient mismatch occurred and the Freedom SOLO showed superior durability.

Keywords: Aortic valve bioprosthesis; Aortic valve replacement; Small aortic root; Stentless bioprosthesis.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Bioprosthesis*
  • Disease-Free Survival
  • Echocardiography
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Netherlands
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies
  • Risk Factors
  • Stents*
  • Time Factors
  • Treatment Outcome