9-Month Clinical and Angiographic Outcomes of the COBRA Polyzene-F NanoCoated Coronary Stent System

JACC Cardiovasc Interv. 2017 Jan 23;10(2):160-167. doi: 10.1016/j.jcin.2016.10.037.

Abstract

Objectives: The aim of this study was to assess the safety and effectiveness of the COBRA Polyzene-F NanoCoated Coronary Stent System (CeloNova Biosciences, San Antonio, Texas) for the treatment of de novo coronary artery lesions.

Background: Polyzene-F-coated coronary stents have shown reduced thrombogenicity and inflammation in preclinical studies.

Methods: Patients with de novo coronary artery lesions meeting eligibility criteria were enrolled in a nonrandomized, prospective clinical trial. The primary endpoint was target vessel failure (TVF) (defined as a composite of cardiac death, myocardial infarction, or clinically driven target vessel revascularization) at 9 months. A pre-specified subset was planned for routine repeat angiographic follow-up at 9 months. The powered secondary endpoint was mean late lumen loss (LL). The comparator was a performance goal derived from meta-analysis of historical bare-metal stent trials of 19.62% for TVF and 1.1 mm for LL. Other secondary endpoints were clinically driven target lesion revascularization and definite or probable stent thrombosis.

Results: Of 296 enrolled patients, 287 (97%) completed primary endpoint analysis; 130 were planned for angiographic follow-up and 115 (88%) completed. At 9 months, TVF had occurred in 33 patients (11.5%; upper 95% confidence boundary: 15.07%), including 1 (0.3%) cardiac death, 20 (7.0%) myocardial infarctions (17 periprocedural), and 17 (5.9%) target vessel revascularizations. LL was 0.84 ± 0.48 mm (upper 95% confidence boundary: 0.92). Target lesion revascularization occurred in 13 patients (4.6%). There were no stent thrombosis events.

Conclusions: The COBRA Polyzene-F stent met performance goals for TVF and LL at 9 months. There was an excellent safety profile, with infrequent late myocardial infarction and no stent thrombosis.

Keywords: restenosis; stent(s); thrombosis.

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Coated Materials, Biocompatible*
  • Coronary Angiography*
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / therapy*
  • Coronary Restenosis / diagnostic imaging*
  • Coronary Restenosis / etiology
  • Coronary Thrombosis / diagnostic imaging
  • Coronary Thrombosis / etiology
  • Coronary Vessels / diagnostic imaging*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Nanoparticles*
  • Organophosphorus Compounds / chemistry
  • Polymers / chemistry*
  • Predictive Value of Tests
  • Prospective Studies
  • Prosthesis Design
  • Risk Factors
  • Stents*
  • Time Factors
  • Treatment Outcome

Substances

  • Coated Materials, Biocompatible
  • Organophosphorus Compounds
  • Polymers