Left Ventricular Outflow Tract Migration of a Balloon-Expandable Prosthesis During Transcatheter Aortic Valve Implantation

Int Heart J. 2017 Apr 6;58(2):290-293. doi: 10.1536/ihj.16-288. Epub 2017 Mar 17.

Abstract

Valve migration into the left ventricular outflow tract (LVOT) during transcatheter aortic valve implantation (TAVI) is a life-threatening complication. An 89-year-old female patient was admitted for TAVI due to severe symptomatic aortic stenosis. After deployment of a balloon-expandable prosthesis, the prosthesis had migrated into the LVOT. The prosthesis was reimpacted to the aortic annulus by a balloon-assisted recapture procedure. Immediately after recapturing the prosthesis with an oversized balloon, the patient's vital signs deteriorated due to acute aortic regurgitation (AR), and a prompt valve-in-valve (V-in-V) procedure allowed us to stabilize the patient's condition. This is the first reported case of a V-in-V procedure using an oversized balloon and a larger prosthesis to treat migration of the initial prosthesis into the LVOT. Balloon recapture and V-in-V procedure using an oversized balloon and larger prosthesis for a migrated balloonexpandable prosthesis into the LVOT is feasible, but hemodynamic support should be prepared before recapture and Vin-V because overdilatation of the first prosthesis might cause hemodynamic collapse due to severe AR.

Publication types

  • Case Reports

MeSH terms

  • Aged, 80 and over
  • Aortic Valve
  • Female
  • Heart Valve Prosthesis
  • Humans
  • Prosthesis Failure
  • Transcatheter Aortic Valve Replacement / adverse effects*