Automation of sample processing for ICP-MS determination of 90Sr radionuclide at ppq level for nuclear technology and environmental purposes

Talanta. 2017 Jul 1:169:216-226. doi: 10.1016/j.talanta.2016.10.051. Epub 2016 Oct 13.

Abstract

90Sr is a widely determined radionuclide for environmental purposes, nuclear waste control, and can be also monitored in coolants in nuclear reactor plants. In the developed method, the ICP-MS detection was employed together with sample processing in sequential injection analysis (SIA) setup, equipped with a lab-on-valve with mechanized renewal of sorbent bed for solid-phase extraction. The optimized conditions of determination included preconcentration of 90Sr on cation-exchange column and removal of different type of interferences using extraction Sr-resin. The limit of detection of the developed procedure depends essentially on the configuration of the employed ICP-MS spectrometer and on the available volume of the sample to be analyzed. For 1L initial sample volume, the method detection limit (MDL) value was evaluated as 2.9ppq (14.5BqL-1). The developed method was applied to analyze spiked river water samples, water reference materials, and also simulated and real samples of the nuclear reactor coolant.

Keywords: Environmental analysis; Extraction resin; ICP-MS; Nuclear reactor coolant; Solid-phase extraction; Strontium-90.