Long-term outcomes of a rapid deployment aortic valve: data up to 5 years

Eur J Cardiothorac Surg. 2017 Aug 1;52(2):281-287. doi: 10.1093/ejcts/ezx103.

Abstract

Objectives: Rapid deployment aortic valve replacement (AVR) has been developed to facilitate minimally invasive approaches for cardiac surgery and shorten procedural times. TRITON is a prospectively designed study to assess safety and efficacy of rapid deployment AVR with Edwards INTUITY valve system. This report presents the 5-year outcomes of the TRITON trial.

Methods: A total of 295 patients with aortic valve stenosis were enrolled in the TRITON trial and 287 patients received the study valve. Procedural, early (≤30 days) and late (>30 days) outcomes were collected. Valve haemodynamic performance was evaluated at specified time points by an independent Echocardiography Core Laboratory and clinical events adjudicated by an independent Clinical Events Committee.

Results: Mean age was 75.3 ± 6.7 years. A total of 158 patients underwent isolated AVR and 129 patients underwent AVR with concomitant procedures. The 5-year survival rates were 85.7 ± 3.4% and 75.2 ± 4.9% for isolated AVR and concomitant AVR, respectively. Overall, freedom from valve-related death at 5 years was 98.2 ± 0.8%. At 5 years, the valve effective orifice area was 1.6 ± 0.3 cm 2 , mean gradient was 10.5 ± 5.4 mmHg and peak gradient was 18.9 ± 9.3 mmHg.

Conclusions: The 5-year outcomes of the TRITON trial demonstrate acceptable long-term safety and excellent haemodynamic performance of rapid deployment AVR with the Edwards INTUITY valve system.

Clinicaltrials.gov: NCT01445171.

Keywords: Aortic valve replacement; Bioprosthesis; Haemodynamics; Heart valve.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / surgery*
  • Female
  • Heart Valve Prosthesis Implantation / instrumentation
  • Heart Valve Prosthesis Implantation / methods
  • Heart Valve Prosthesis Implantation / mortality*
  • Heart Valve Prosthesis Implantation / statistics & numerical data*
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Prospective Studies
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01445171