Background: Information regarding suitability for subcutaneous defibrillator (sICD) implantation in tetralogy of Fallot (ToF) and systemic right ventricle is scarce and needs to be further explored. The main objective of our study was to determine the proportion of patients with ToF and systemic right ventricle eligible for sICD with both, standard and right-sided screening methods. Secondary objectives were: (i) to study sICD eligibility specifically in patients at high risk of sudden cardiac death, (ii) to identify independent predictors for sICD eligibility, and (iii) to compare the proportion of eligible patients in a nonselected ICD population.

Methods: We recruited 102 patients with ToF, 33 with systemic right ventricle, and 40 consecutive nonselected patients. Conventional electrocardiographic screening was performed as usual. Right-sided alternative screening was studied by positioning the left-arm and right-arm electrodes 1 cm right lateral of the xiphoid midline. The Boston Scientific ECG screening tool was utilized.

Results: In high-risk patients with ToF, eligibility was higher with right-sided screening in comparison with standard screening (61% vs. 44%; p = .018). Eligibility in high-risk right ventricle population was identical with both screening methods (77%, p = ns). The only independent predictor for sICD eligibility was QRS duration.

Conclusion: In high-risk patients with ToF, right-sided implantation of the sICD could be an alternative to a conventional ICD. In patients with a systemic right ventricle, implantation of a sICD is an alternative to a conventional sICD.

Keywords: ECG screening; right-sided screening; subcutaneous implantable cardioverter-defibrillator; systemic right ventricle; tetralogy of Fallot.

MeSH terms

  • Adult
  • Death, Sudden, Cardiac / prevention & control*
  • Defibrillators, Implantable*
  • Electrocardiography / methods*
  • Electrocardiography / statistics & numerical data
  • Female
  • Heart Ventricles / physiopathology
  • Humans
  • Male
  • Middle Aged
  • Patient Selection*
  • Risk Factors
  • Tetralogy of Fallot / physiopathology*
  • Tetralogy of Fallot / therapy