Serum Metal Levels for Diagnosis of Adverse Local Tissue Reactions Secondary to Corrosion in Metal-on-Polyethylene Total Hip Arthroplasty

J Arthroplasty. 2017 Sep;32(9S):S272-S277. doi: 10.1016/j.arth.2017.04.016. Epub 2017 Apr 20.

Abstract

Background: Recently, corrosion at the head-neck junction in metal-on-polyethylene bearing surface total hip arthroplasty (THA) has been recognized as a cause of adverse local tissue reactions (ALTRs). Serum metal levels have been advocated as a tool for the diagnosis of ALTR; however, no prior studies have specifically examined their utility. The purpose of this study was to determine the optimal cutoff values for serum cobalt and chromium levels in diagnosing ALTR after metal-on-polyethylene bearing surface THA.

Methods: We reviewed 447 consecutive patients with serum metal levels tested at our institution and identified 64 patients with a metal-on-polyethylene bearing who had axial imaging or underwent reoperation to confirm the presence or absence of ALTR. Receiver-operating characteristic curves were produced to identify cutoff thresholds to optimize sensitivity, and diagnostic test performance was characterized.

Results: Forty-four of the 64 patients (69%) were positive for an ALTR. The best test for the diagnosis of ALTR was the serum cobalt level (area under the curve [AUC] = 99%). A threshold cutoff of ≥1.0 ng/mL had a sensitivity of 100%, specificity of 90%, positive predictive value (PPV) of 96%, and negative predictive value (NPV) of 100%. Serum chromium levels were also diagnostic (AUC = 87%). A threshold cutoff of ≥0.15 ng/mL had a sensitivity of 100%, specificity of 50%, PPV of 81%, and NPV of 100%. Finally, serum cobalt-to-chromium ratio was also helpful for diagnosis (AUC = 90%). A threshold cutoff value of 1.4 for the cobalt-to-chromium ratio offered a sensitivity of 93%, specificity of 70%, PPV of 87%, and NPV of 82%.

Conclusion: Measurement of serum cobalt level with a threshold value of 1.0 ng/mL in our experience is the best test for identifying the presence of ALTR in patients with a metal-on-polyethylene THA. Measurement of chromium level and the ratio of cobalt-to-chromium levels are also of value.

Keywords: adverse local tissue reaction; mechanically assisted crevice corrosion; metal-on-polyethylene total hip arthroplasty; modular junction corrosion; serum metal levels.

MeSH terms

  • Adult
  • Aged
  • Area Under Curve
  • Arthroplasty, Replacement, Hip / adverse effects*
  • Chromium / blood*
  • Cobalt / blood*
  • Corrosion
  • Female
  • Hip Prosthesis / adverse effects*
  • Humans
  • Male
  • Metals / blood*
  • Middle Aged
  • Polyethylene / chemistry*
  • Predictive Value of Tests
  • Prosthesis Design
  • Prosthesis Failure*
  • ROC Curve
  • Reoperation
  • Sensitivity and Specificity
  • Treatment Outcome