A worldwide experience of the management of battery failures and chronic device retrieval of the Nanostim leadless pacemaker

Heart Rhythm. 2017 Dec;14(12):1756-1763. doi: 10.1016/j.hrthm.2017.07.004. Epub 2017 Jul 10.

Abstract

Background: The Nanostim leadless pacemaker (LP) met the primary endpoints in an investigational device exemption trial, and was shown to be fully retrievable percutaneously. In October 2016, St Jude Medical issued a worldwide alert of a battery malfunction that caused lost pacing output and LP communication.

Objective: To report the battery failure mechanism and incidence and the worldwide patient management, including device retrieval experiences.

Methods: The affected LP battery is a custom lithium-carbon monofluoride cell. These were returned after failure and underwent analysis assessing electronics and battery performance. Data were collected in ongoing clinical studies when LPs were abandoned or retrieved.

Results: Of 1423 LPs implanted worldwide, there were 34 battery failures, occurring at 2.9 ± 0.4 years with no instances of associated patient injury. Analysis of returned batteries revealed an increase in battery resistance caused by insufficient electrolyte availability at the cathode/anode interface. A total of 66 of 73 retrieval attempts were successful (90.4%; implant duration range: 0.2-4.0 years). The LP docking button was inaccessible in 6 patients, and the docking button detached from the LP during retrieval in 1 patient. There was 1 case of arteriovenous fistula and another case of the LP docking button migrating into the pulmonary artery. There were also 115 non-LP retrieval patients after the advisory who received an additional pacemaker, with no adverse device-to-device interactions reported.

Conclusion: As with standard pacers, LPs can have critical battery failures. Chronic retrieval of LPs is safe and efficacious.

Keywords: Battery advisory; Battery failure; Extraction; Leadless pacemaker; Patient management; Retrieval.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Arrhythmias, Cardiac / therapy*
  • Device Removal / methods*
  • Disease Management*
  • Equipment Design
  • Equipment Failure
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Pacemaker, Artificial / adverse effects*
  • Patient Safety
  • Retrospective Studies
  • Time Factors