Investigating the Robustness Neighborhood Gray Tone Difference Matrix and Gray Level Co-occurrence Matrix Radiomic Features on Clinical Computed Tomography Systems Using Anthropomorphic Phantoms: Evidence From a Multivendor Study

J Comput Assist Tomogr. 2017 Nov/Dec;41(6):995-1001. doi: 10.1097/RCT.0000000000000632.

Abstract

Objective: The aim of this study was to determine if optimized imaging protocols across multiple computed tomography (CT) vendors could result in reproducible radiomic features calculated from an anthropomorphic phantom.

Methods: Materials with varying degrees of heterogeneity were placed throughout the lungs of the phantom. Twenty scans of the phantom were acquired on 3 CT manufacturers with chest CT protocols that had optimized protocol parameters. Scans were reconstructed using vendor-specific standards and lung kernels. The concordance correlation coefficient (CCC) was used to calculate reproducibility between features. For features with high CCC values, Bland-Altman analysis was also used to quantify agreement.

Results: The mean Hounsfield unit (HU) was 32.93 HU (141.7 to -26.5 HU) for the rubber insert and 347.2 HU (-320.9 to -347.7 HU) for the wood insert. Low CCC values of less than 0.9 were calculated for all features across all scans.

Conclusions: Radiomic features that are derived from the spatial distribution of voxel intensities should be particularly scrutinized for reproducibility in a multivendor environment.

Publication types

  • Comparative Study

MeSH terms

  • Humans
  • Lung
  • Phantoms, Imaging*
  • Reproducibility of Results
  • Tomography Scanners, X-Ray Computed*
  • Tomography, X-Ray Computed*