Closure of Secundum Atrial Septal Defects With the AMPLATZER Septal Occluder: A Prospective, Multicenter, Post-Approval Study

Circ Cardiovasc Interv. 2017 Aug;10(8):e004212. doi: 10.1161/CIRCINTERVENTIONS.116.004212.

Abstract

Background: Prospective data on the medium-term safety and effectiveness of the AMPLATZER Septal Occluder in clinical practice are not available. The objective of this study was to prospectively evaluate the risk of hemodynamic compromise and obtain medium-term survival data on patients implanted with the AMPLATZER Septal Occluder for percutaneous closure of secundum atrial septal defects.

Methods and results: Subjects were enrolled prospectively at 50 US sites and followed for 2 years. Between 2008 and 2012, atrial septal defect closure with the AMPLATZER Septal Occluder was attempted in 1000 patients (aged 0.3-83.6 years, mean 21±22 years). Procedural closure occurred in 97.9%, with 1-month and 2-year closure 98.5% and 97.9%, respectively. Hemodynamic compromise occurred in 6 subjects (0.65%), because of dysrhythmia in 2, device embolization in 1, and cardiac erosion in 3. The rate of cardiac erosion was 0.3% (average 83, range 12-171 days from implant).

Conclusions: Closure of atrial septal defect with the AMPLATZER Septal Occluder is safe and effective. The rate of hemodynamic compromise and cardiac erosion is rare. The risk factors for cardiac erosion after device closure are not yet clear.

Clinical trial registration: URL: http://www.clinicaltrials.gov. Unique identifier: NCT00650936.

Keywords: catheterization; echocardiography; heart; heart septal defects; risk factors.

Publication types

  • Clinical Study
  • Multicenter Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Child
  • Child, Preschool
  • Female
  • Heart Septal Defects, Atrial / therapy*
  • Humans
  • Infant
  • Male
  • Middle Aged
  • Prospective Studies
  • Septal Occluder Device / adverse effects*
  • Septal Occluder Device / statistics & numerical data
  • Young Adult

Supplementary concepts

  • Atrial Septal Defect, Secundum Type

Associated data

  • ClinicalTrials.gov/NCT00650936