Clinical outcomes of AF patients treated with the first and second-generation of circular mapping and ablation catheter: insights from a real world multicenter experience

J Interv Card Electrophysiol. 2017 Dec;50(3):245-251. doi: 10.1007/s10840-017-0278-y. Epub 2017 Sep 5.

Abstract

Purpose: Complete pulmonary vein isolation (PVI) is the best documented target for catheter ablation, and different technologies have shown comparable outcomes. The multielectrode phased-RF/duty cycled (PhRF/DC) pulmonary vein ablation catheter (PVAC) and its second generation (PVAC-GOLD) have shown promising clinical results in single and multicenter experiences. Our aim is to assess and compare the safety and efficacy in the real clinical practice among two generations of circular PhRF/DC catheters by performing PVI in patients suffering from recurrent atrial fibrillation (AF).

Methods: Eighty-four AF patients treated with PVAC and 64 with PVAC-GOLD were prospectively followed in five Italian cardiology centers in the mainframe of the 1STOP-ClinicalService project.

Results: Fluoroscopic and total procedure time were significantly different in the two groups. In particular, in the PVAC-GOLD group, the mean fluoroscopic time was 22.8 ± 12.7 min vs 31.6 ± 18.9 in the PVAC group (p = 0.002), and the mean total procedure duration was 117.6 ± 36.0 vs 147.4 ± 40.6, in the PVAC-GOLD group and the PVAC group, respectively (p = 0.001). Only two out of 148 patients reported a peri-procedural complication. Over 20.9 ± 12.0 months of follow-up, AF recurrence occurred in 58 patients. Kaplan-Meier freedom from AF recurrence did not differ between the two groups (64.1 ± 10% in the PVAC group vs 68.2 ± 9% in the PVAC-GOLD group at 1 year, p = ns).

Conclusions: In our multicenter analysis, AF ablation using two generations of circular PhRF/DC catheters is safe and effective. No difference was observed in terms of safety and efficacy of the AF ablation between the two catheters, with the mean procedural time being shorter in the PVAC-GOLD group.

Keywords: Ablation of atrial fibrillation; Atrial fibrillation; PVAC-GOLD; Pulmonary vein isolation.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Atrial Fibrillation / diagnostic imaging*
  • Atrial Fibrillation / mortality
  • Atrial Fibrillation / surgery*
  • Body Surface Potential Mapping / instrumentation*
  • Cardiac Catheters
  • Catheter Ablation / methods*
  • Catheter Ablation / mortality
  • Cohort Studies
  • Equipment Design
  • Equipment Safety
  • Female
  • Fluoroscopy / methods
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Operative Time
  • Pulmonary Veins / diagnostic imaging
  • Pulmonary Veins / surgery*
  • Recurrence
  • Retrospective Studies
  • Risk Assessment
  • Severity of Illness Index
  • Survival Rate
  • Time Factors
  • Treatment Outcome