Evaluation of potential substrates for restenosis and thrombosis in overlapped versus edge-to-edge juxtaposed bioabsorbable scaffolds: Insights from a computed fluid dynamic study

Cardiovasc Revasc Med. 2018 Apr;19(3 Pt A):273-278. doi: 10.1016/j.carrev.2017.08.005. Epub 2017 Aug 17.

Abstract

Background/purpose: Multiple BRSs and specifically the Absorb scaffold (BVS) (Abbott Vascular, Santa Clara, CA USA) have been often used to treat long diffuse coronary artery lesions. We evaluate by a computational fluid dynamic(CFD) study the impact on the intravascular fluid rheology on multiple bioabsorbable scaffolds (BRS) by standard overlapping versus edge-to-edge technique.

Methods/materials: We simulated the treatment of a real long significant coronary lesion (>70% luminal narrowing) involving the left anterior descending artery (LAD) treated with a standard or edge-to-edge technique, respectively. Simulations were performed after BVS implantations in two different conditions: 1) Edge-to-edge technique, where the scaffolds are kissed but not overlapped resulting in a luminal encroachment of 0.015cm (150μm); 2) Standard overlapping, where the scaffolds are overlapped resulting in a luminal encroachment of 0.030cm (300μm). After positioning the BVS across the long lesion, the implantation procedure was performed in-silico following all the usual procedural steps.

Results: Analysis of the wall shear stress (WSS) suggested that at the vessel wall level the WSS were lower in the overlapping zones overlapping compared to the edge-to-edge zone (∆=0.061Pa, p=0.01). At the struts level the difference between the two WSS was more striking (∆=1.065e-004 p=0.01) favouring the edge-to-edge zone.

Conclusions: Our study suggested that at both vessel wall and scaffold struts levels, there was lowering WSS when multiple BVS were implanted with the standard overlapping technique compared to the "edge-to-edge" technique. This lower WSS might represent a substrate for restenosis, early and late BVS thrombosis, potentially explaining at least in part the recent evidences of devices poor performance.

Keywords: BVS; BVS restenosis; BVS thrombosis; CFD.

MeSH terms

  • Absorbable Implants*
  • Blood Flow Velocity
  • Computer Simulation*
  • Coronary Angiography
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / physiopathology
  • Coronary Artery Disease / surgery*
  • Coronary Circulation
  • Coronary Restenosis / diagnostic imaging
  • Coronary Restenosis / etiology*
  • Coronary Restenosis / physiopathology
  • Coronary Stenosis / diagnostic imaging
  • Coronary Stenosis / physiopathology
  • Coronary Stenosis / surgery*
  • Coronary Thrombosis / diagnostic imaging
  • Coronary Thrombosis / etiology*
  • Coronary Thrombosis / physiopathology
  • Humans
  • Models, Cardiovascular*
  • Percutaneous Coronary Intervention / adverse effects*
  • Percutaneous Coronary Intervention / instrumentation*
  • Prosthesis Design
  • Rheology
  • Risk Factors
  • Stents*
  • Treatment Outcome