Objective: Aim of this study was to assess the artifact reduction in patients with orthopedic hardware in the spine as provided by (1) metal-artifact-reduction algorithms (O-MAR) and (2) virtual monoenergetic images (MonoE) as provided by spectral detector CT (SDCT) compared to conventional iterative reconstruction (CI).
Methods: In all, 28 consecutive patients with orthopedic hardware in the spine who underwent SDCT-examinations were included. CI, O-MAR and MonoE (40-200 keV) images were reconstructed. Attenuation (HU) and noise (SD) were measured in order to calculate signal-to-noise ratio (SNR) of paravertebral muscle and spinal canal. Subjective image quality was assessed by two radiologists in terms of image quality and extent of artifact reduction.
Results: O-MAR and high-keV MonoE showed significant decrease of hypodense artifacts in terms of higher attenuation as compared to CI (CI vs O-MAR, 200 keV MonoE: -396.5HU vs. -115.2HU, -48.1HU; both p ≤ 0.001). Further, artifacts as depicted by noise were reduced in O-MAR and high-keV MonoE as compared to CI in (1) paravertebral muscle and (2) spinal canal-CI vs. O-MAR/200 keV: (1) 34.7 ± 19.0 HU vs. 26.4 ± 14.4 HU, p ≤ 0.05/27.4 ± 16.1, n.s.; (2) 103.4 ± 61.3 HU vs. 72.6 ± 62.6 HU/60.9 ± 40.1 HU, both p ≤ 0.001. Subjectively both O-MAR and high-keV images yielded an artifact reduction in up to 24/28 patients.
Conclusion: Both, O-MAR and high-keV MonoE reconstructions as provided by SDCT lead to objective and subjective artifact reduction, thus the combination of O-MAR and MonoE seems promising for further reduction.
Keywords: Metal artifact reduction; Orthopedic hardware; Spectral detector computed tomography; Spinal instrumentation; Virtual monoenergetic images.