A novel guidewire-integrated embolic protection filter device with a handy-folding system: In vitro and in vivo performance assessment

Catheter Cardiovasc Interv. 2018 Jul;92(1):E9-E14. doi: 10.1002/ccd.27315. Epub 2017 Sep 20.

Abstract

Objectives: We developed a novel guidewire-integrated filter device with a handy-folding system (IFD: Inoue filter device). In vitro and in vivo studies were conducted to evaluate the feasibilityof the IFD.

Background: Although distal atheromatous and thrombotic embolizations remain unresolved critical issues during catheter interventions, distal protection devices are infrequently used partly because of reduced lower maneuverability.

Methods: In the in vitro experiment, we created an experimental circulation model composed of silicone latex tubes, a reservoir, and a roller pump. After the filter device was deployed in the tube, polystyrene fluorescent microspheres were injected and the capture rate was calculated. Ten trials were performed using the IFD and Spider FX. In the in vivo study, five independent operators deployed, and they retrieved the IFD in swine common iliac and internal carotid arteries. The procedural success rate as well as the delivery and retrieval time was evaluated.

Results: In the in vitro study, the mean capture rate was 94% and 35% in the IFD and Spider groups, respectively. In the in vivo study, all procedures were successful, with no complications. The mean delivery time was 281 ± 87 s and 194 ± 67 s and the mean retrieval time was 24 ± 9 and 13 ±1 s in the left internal carotid and the left common iliac arteries, respectively.

Conclusion: Although further studies and improvements are required, the study results indicate that the IFD is feasible.

Keywords: carotid artery disease; embolic protection devices; intervention; percutaneous coronary intervention; peripheral intervention.

Publication types

  • Video-Audio Media

MeSH terms

  • Angiography
  • Animals
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Carotid Artery, Internal* / diagnostic imaging
  • Catheterization, Peripheral / adverse effects
  • Catheterization, Peripheral / instrumentation*
  • Device Removal
  • Embolic Protection Devices*
  • Iliac Artery* / diagnostic imaging
  • Materials Testing
  • Models, Anatomic
  • Models, Animal
  • Models, Cardiovascular
  • Prosthesis Design
  • Sus scrofa
  • Time Factors
  • Vascular Access Devices*