Multicenter, multivendor phantom study to validate proton density fat fraction and T2* values calculated using vendor-provided 6-point DIXON methods

Clin Imaging. 2018 Sep-Oct:51:38-42. doi: 10.1016/j.clinimag.2018.01.011. Epub 2018 Feb 2.

Abstract

Purpose: To evaluate the reproducibility of proton density fat fraction (PDFF) and T2* in a fat-water phantom on three different 3 T MRI systems using 6-point DIXON methods.

Methods: A phantom which included varying fat volume percentages (true fat fraction [FF]) was scanned by three 3 T MR machines, and PDFF and T2* were measured.

Results: The mean difference between true FF and PDFF was small in all vendors (-2.11% to 0.41%). However, the difference ratio for T2* values was large among vendors (1.79 to 3.36).

Conclusions: True FF and PDFF were consistent across vendors; however, T2* varied greatly.

Keywords: DIXON; Fat; Liver; MRI; Phantom; Quantification.